BIPAP ST


Includes ST (Spontaneous Timed) feature. The feature of this device is to allow the patient to breathe spontaneously, but the respiratory frequency can be adjusted by the device. BİPAP ST device should be used in cases where breathing is not regulated by BİPAP, higher pressure treatment is required, and especially in the presence of Central Apneas.

In addition to the diagnosis of Obstructive Sleep Apnea Syndrome; It is recommended to be used in patients with a pathology that causes alveolar hypoventilation such as Restrictive Pulmonary Disease, COPD, Obesity-Hypoventilation Syndrome, Sleep-Related Hypoventilation/Hypoxemic Syndrome, after determining spontaneous breathing and insufficient triggering power.

While the breathing frequency is dependent on the person in BİPAP devices, BİPAP ST devices trigger breathing in a way that provides the defined number of breathing.

As long as the patient breathes spontaneously, the device operates in the "BİPAP S" mode. When breathing is stopped, the device automatically switches to the "BİPAP T" mode, that is, the forced breathing mode.


Technical Specifications

IPAP

3 - 35 mbar (up to 35.0 mbar in 0.1 mbar increments) Breathing.

EPAP

3 - 35 mbar (up to 35.0 mbar in 0.1 mbar increments) Exhaling.

RAMP TIME

OFF / 5-10-15...90 MIN.

RAMP PRESSURE

It can be adjusted between 3 - 35 mbar. It gradually rises to the IPAP / EPAP value.

TI

Inspiratory time (0.1-0.2-0.3...7.5 sec.)

I-SLOPE

OFF / 1-2-3-4-5 Length/rate of patient's breathing time (Time taken by the device to go from EPAP (positive exhalation pressure level) to IPAP (positive exhalation pressure level.).

E-SLOPE

OFF / 1-2-3-4-5 Length/rate of the patient's exhalation time (Time taken by the device to descend from IPAP (positive inspiratory pressure level) to EPAP (positive exhalation pressure level).

I SENSITIVITY

OFF / 1-2-3-4-5 It is the sensitivity to detect inspiration (the sensitivity to detect the beginning of the breathing phase of breathing.).

E SENSITIVITY

OFF / 1-2-3-4-5 It is the sensitivity to detect expiration (the sensitivity to detect the beginning of the exhalation phase of breathing.).

BPM-FREQUENCY-RESPIRATORY NUMBER

5…35 Breaths (The number of breaths per minute.)

TUBE DIAMETER

12mm/15mm/22 mm

SIGH

OFF / 300-350-400-450…1000 In one breath, it sends some extra air to the patient, allowing him to take a deep breath. The device uses this feature if the patient encounters conditions such as Apnea/Hypoapnea/Central Apnea duringtherapy. This feature is not suitable for use in patients with COPD.

PRESSURE UNIT

(hPa/mbar/cmH2o)